Description of the Upgrade
Installation Procedure for the Upgrade
Step 23. Install the Front Panel Overlay
To see an image of the front panel overlay, click the Chapter 6 bookmark "Front
Panel Assembly, Front Side, All Options" in the PDF Service Guide
are listed in
1. Remove the protective backing from the new front panel overlay
2. Loosely place the overlay in the recess on the front panel.
3. Placing two fingers at the middle, press the overlay firmly onto the frame
Step 24.Install the Front Panel Jumper Cables
For instructions on installing the W30 front panel jumpers (N5222-20091),
click the Chapter 7 bookmark "Removing and Replacing the Front Panel
Assembly" in the PDF Service Guide
Step 25. Position the Cables and Wires to Prevent Pinching
On the top side of the PNA, carefully position the gray flex cables so they can't
be pinched between the covers and the rails.
On the bottom side of the PNA, carefully fold or push down the ribbon cables
and wires so they can't be pinched between the hardware and the outer cover.
Ribbon cables and wires must never be positioned on top of hardware.
Step 26. Reinstall the Inner Cover
For instructions, click the Chapter 7 bookmark "Removing the Covers" in the
PDF Service Guide
Step 27. Reinstall the Outer Cover
For instructions, click the Chapter 7 bookmark "Removing the Covers" in the
PDF Service Guide
Step 28. Remove Option 217 License
Procedure Requirements
— The analyzer must be powered up and operating to perform this procedure.
— The Network Analyzer program must not be running.
— A keyboard and mouse must be connected to the network analyzer.
44
Table 1-1 on page
(N5242-80031 or N5242-80034).
while sliding your fingers in opposite directions towards the ends of the
overlay. Repeat on all areas of the overlay.
1
.
1
.
1. See
"Downloading the Online PNA–X Service Guide" on page
13.
1
.
Installation Note N5242-90127
1
. New parts
11.