Description of the Upgrade
Installation Procedure for the Upgrade
Step 17. Install the A8 26.5 GHz Source 2 Assembly
'Refer to
listed in
Install the A8 26.5 GHz source 2 board (5087-7837) in the analyzer as shown.
Secure board with two screws (item
Figure 14
A8 Source 2 Assembly Installation (0515-0380 (x2))
Installation Note N5242-90127
Figure 14
for this part of this step of the procedure. New parts are
Table 1-1 on page
13.
; 0515-0380) and torque to 21 in-lbs.
33