Description of the Upgrade
Installation Procedure for the Upgrade
Step 19.Install the Test Set Cables
Follow instructions carefully when making cable connections, especially
wire harness connections. Incorrect connections can destroy components,
resulting in additional customer costs.
Be careful not to damage the center pins of the semi-rigid cables. Some
flexing of the cables may be necessary but do not over-bend them.
Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections
except the front and rear panel cable connectors. Torque these
connections to 21 in-lb.
Cables that are to be reinstalled are designated with "reuse." Part numbers
in a normal weight font are for analyzers with serial numbers
MY/SG/US5310 and below. Part numbers in a bold font are for analyzers
with serial numbers MY/SG/US5321 and above.
Install the following semi-rigid cables in the order listed. To see images
showing the location of these cables, click the Chapter 6 bookmark "Bottom
RF Cables, 4-Port, Option 417, S/N Prefix ≥6021" in the PDF Service Guide
New parts are listed in
— W26 (N5222-20034) A32 port 2 coupler to front-panel Port 2 CPLR ARM
— W93 (reuse) (N5242-20047 or N5222-20037) Port 2 RCVR B IN to A45 port
— W84 (reuse) (N5242-20046 or N5222-20036) A37 port 2 source attenuator
— W135 (reuse) (Opt 029A only) (N5242-20073 or N5242-20278) Front-panel
— W128 (reuse) (Opt 029A only) (N5242-20134 or N5242-20303) A53 port 2
— W127 (reuse) (Opt 029A only) (N5242-20116 or N5242-20292) A37 port 2
— W130 (reuse) (Opt 029A only) (N5242-20133) A53 port 2 bypass switch to
— W129 (reuse) (Opt 029A only) (N5242-20117 or N5242-20293) A53 port 2
38
Table 1-1 on page
For all analyzer serial numbers:
2 receiver attenuator
to front-panel Port 2 SOURCE OUT
Port 2 RCVR B IN to A45 port 2 receiver attenuator
bypass switch to front-panel Port 2 SOURCE OUT
source attenuator to A53 port 2 bypass switch
A54 port 2 bridge
bypass switch to A54 port 2 bridge
For all analyzer serial numbers:
1. See
"Downloading the Online PNA–X Service Guide" on page
13.
11.
Installation Note N5242-90127
1
.