Description of the Upgrade
Installation Procedure for the Upgrade
B. Remove the DDS Assembly
Follow the two instructions shown in
The (x2) DDS assembly will not be reused.
Figure 12
Remove the Old (x2) DDS Assembly
Step 16. Assemble the Source 2 Board
Install Cables on Source 2 Board
Refer to
listed in
1. Attach new cables W5 (N5222-20062) and W7 (N5222-20063) to the A8
2. Use a 5/16-in torque wrench set to 10 in-lbs to tighten the cable
Installation Note N5242-90127
Figure 13
of this document for this part of this step. New parts are
Table 1-1 on page 13
26.5 GHz source 2 board as shown. Make sure that both cables are parallel
to the A8 26.5 GHz source board as shown in the END VIEW. Cable W5 is
the longer of the two cables.
connectors.
Figure
12.
of this document.