Cleaning And Disinfecting; Maintenance, Care And Service - Microlife NEB 200 Manual

Compressor nebuliser
Hide thumbs Also See for NEB 200:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
3. Connect the nebuliser with the air tube 6 to the compressor 1
and plug the power lead 2 into the socket (230V 50 Hz AC).
4. To start the treatment, set ON/OFF switch 3 into the «I»
position.
 The mouthpiece gives you a better drug delivery to the
lungs.
 Choose between adult 8 or child face mask 9 and make
sure that it encloses the mouth and nose area completely.
 Use all accessories including the nose piece AL as
prescribed by your doctor.
5. During inhalation, sit upright and relaxed at a table and not in an
armchair, in order to avoid compressing your respiratory
airways and impairing the treatment effectiveness. Do not lie
down while inhaling. Stop inhalation if you feel unwell.
6. After completing the inhalation period recommended by your
doctor, switch the ON/OFF switch 3 to position «O» to turn off
the device and unplug it from the socket.
7. Empty the remaining medication from the nebuliser and clean
the device as described in the section «Cleaning and disin-
fecting»
This device was designed for intermittent use of
30 min. On / 30 min. Off. Switch off the device after 30 min.
use and wait for another 30 min. before you resume
treatment.
The device requires no calibration.
No modification to the device is permitted.

3. Cleaning and disinfecting

Thoroughly clean all components to remove medication residuals
and possible impurities after each treatment.
Use a soft and dry cloth with non-abrasive cleaners to clean the
compressor.
Make sure that the internal parts of the device are not in
contact with liquids and that the power plug is discon-
nected.
Cleaning and disinfecting of the accessories
Follow carefully the cleaning and disinfecting instructions of the
accessories as they are very important to the performance of the
device and success of the therapy.
Microlife NEB 200
Before and after each treatment
Disassemble the nebuliser 5 by turning the top counterclockwise
and remove the medicine conduction cone. Wash the components
of the disassembled nebuliser, the mouthpiece 7 and the nose
piece AL by using tap water; dip in boiling water for 5 minutes.
Reassemble the nebuliser components and connect it to the air
tube connector, switch the device on and let it work for 10-15
minutes.
Wash masks and air tube with warm water.
Only use cold disinfecting liquids following the
manufacturer's instructions.
Do not boil nor autoclave the air tube and masks.

4. Maintenance, Care and Service

Order all spare parts from your dealer or pharmacist, or contact
Microlife-Service (see foreword).
Replacement of the nebuliser
Replace the nebuliser 5 after a long period of inactivity, in cases
where it shows deformities, breakage, or when the vaporiser head
5-a is obstructed by dry medicine, dust, etc. We recommend to
replace the nebuliser after a period between 6 months and 1 year
depending on the usage.
Only use original nebulisers!
Replacement of the air filter
In normal conditions of use, the air filter AM must be replaced
approximately after 200 working hours or after each year. We
recommend to periodically check the air filter (10 - 12 treatments)
and if the filter shows a grey or brown colour or is wet, replace it.
Extract the filter and replace it with a new one.
Do not try to clean the filter for reusing it.
The air filter shall not be serviced or maintained while in use
with a patient.
Only use original filters! Do not use the device without filter!
EN
3