 Choose between adult 8 or child face mask 9 and make
sure that it encloses the mouth and nose area completely.
 Use all accessories including the nose piece AL as
prescribed by your doctor.
5. During inhalation, sit upright and relaxed at a table and not in an
armchair, in order to avoid compressing your respiratory
airways and impairing the treatment effectiveness. Do not lie
down while inhaling. Stop inhalation if you feel unwell.
6. Inhale the aerosol solution using the prescribed accessory.
7. Once you have finished the treatment, switch the appliance off
by pressing the ON/OFF switch 3.
8. Empty the remaining medication from the nebuliser and clean
the device as described in the section «Cleaning and disin-
fecting»
This device was designed for intermittent use of
30 min. On / 30 min. Off. Switch off the device after 30 min.
use and wait for another 30 min. before you resume
treatment.
The device requires no calibration.
Tampering with the device is absolutely forbidden.
No modification to the device is permitted.

3. Cleaning and disinfecting

Device(s): Aerosoltherapy nebuliser accessories
Before the first use and after each treat-
ment
Follow carefully the cleaning and disinfection
instructions of the accessories as they are
very important for the device performances
and the therapy success.
 Use original accessories only.
 Do not clean or disinfect the air tube.
 The accessories cannot be cleaned and
disinfected by automated method.
 Do not boil nor autoclave the masks.
4
Limitations on
The nebuliser must be replaced after a long
processing
period of inactivity, in case it shows deforma-
tions or breakings, or when the nebulser
nozzle is obstructed by dry medicine, dust,
etc. We recommend to replace the nebuliser
after a period between 6 months and 1 year,
depending on the usage. The maximum times
of cleaning and disinfection of the nebulizer
are 360 times.
The maximum times of cleaning and disinfec-
tion of the mask, the nosepiece and the
mouthpiece are 360 times.
Instructions
Preparation
 Detach the air tube from the nebuliser.
before cleaning
 Detach the mouthpiece or nosepiece or
mask from the nebuliser.
 Make sure that all residual volume is
removed from the nebuliser.
 Disassemble the nebuliser 5 by turning
the top counterclockwise and remove the
medicine conduction cone.
Note: To avoid microbial growth and residual
drug drying, clean and disinfect the accesso-
ries immediately after each use.
Cleaning
Briefly rinse all parts in running tap water
for at least 10 seconds beforehand.
Mix little dish detergent and warm tap water
(i.e. FAIRY brand dish washing liquid for hand
washing in a ratio of 2 ml : 1 L) in a clean
container.
Immerse the components of disassembled
nebulser, the mask, the mouthpiece and the
nosepiece into the mixed water for about 5
min. And then scrub all the surface of all the
components with clean and small brush for at
least 8 times.
Afterwards rinse all parts thoroughly in
running tap water at least 30 seconds, to fully
remove any possible dish washing liquid
residue.