5. Malfunctions and Actions to take

The device cannot be switched on
 Ensure the power lead 2 is correctly plugged into the socket.
 Ensure the ON/OFF switch 3 is in the position «I».
 Make sure that the device has been operating within operating
limits indicated in this manual (30 min On / 30 min Off).
The nebuliser functions poorly or not at all
 Ensure the air tube 6 is correctly connected at both ends.
 Ensure the air tube is not squashed, bent, dirty or blocked. If
necessary, replace with a new one.
 Ensure the nebuliser 5 is fully assembled and the vaporiser
head 5-a is placed correctly and not obstructed.
 Ensure the required medication has been added.
6. Guarantee
This device is covered by a 5 year guarantee from the date of
purchase. During this guarantee period, at our discretion, Microlife
will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
 Transport costs and risks of transport.
 Damage caused by incorrect application or non-compliance
with the instructions for use.
 Damage caused by accident or misuse.
 Packaging/storage material and instructions for use.
 Regular checks and maintenance (calibration).
 Accessories and wearing parts: Nebuliser, masks, mouthpiece,
nose piece, tube, filters, nasal washer (optional).
Should guarantee service be required, please contact the dealer
from where the product was purchased, or your local Microlife
service. You may contact your local Microlife service through our
website:
www.microlife.com/support
Compensation is limited to the value of the product. The guarantee
will be granted if the complete product is returned with the original
invoice. Repair or replacement within guarantee does not prolong
or renew the guarantee period. The legal claims and rights of
consumers are not limited by this guarantee.
4

7. Technical Specifications

Nebulisation rate:
0.4 ml/min. (NaCI 0.9%)
Particle size:
63% < 5 µm
2.83 µm (MMAD)
Max. free air flow:
15 l/min.
Operating air flow:
5.31 l/min.
Acoustic noise level: 52 dBA
Power source:
230V 50 Hz AC
Power lead length:
1.6 m
Nebuliser capacity:
min. 2 ml; max. 8 ml
Residual Volume:
0.8 ml
Operating limits:
30 min. On / 30 min. Off
Operating
10 - 40 °C / 50 - 104 °F
conditions:
10 - 95 % relative maximum humidity
700 - 1060 hPa Atmospheric pressure
Storage and shipping
-20 - +60 °C / -4 - +140 °F
conditions:
10 - 95 % relative maximum humidity
700 - 1060 hPa Atmospheric pressure
Weight:
approx. 1306 g
Dimensions:
103 x 160 x 140 mm
IP Class:
IP21
Reference to
EN 13544-1; EN 60601-1; EN 60601-2;
standards:
EN 60601-1-6; IEC 60601-1-11
Expected service life: 1000 hours
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Class II device as regards protection against electric shocks.
Nebuliser, mouthpiece and masks are type BF applied parts.
The technical specifications may change without
prior notice.