■[F14] Zero cut-off setting
When the pressure display value is close to zero, the product rounds the value and zero will be displayed.
The zero cut-off range can be set in 0.1 [%F.S.] increments between 0 (no zero-cut off) and 10.0 [%F.S.].
<Operation>
Press the
or
Select zero cut-off setting
Press the
or
: The display above is an example for -X578 (2 kPa range) and unit selection function is selected [kPa].
E.g.) 2 kPa range P_1 = 0.03 [kPa], H_1 = 0.02 [kPa], zero cut-off 1.0%
Press the
: When setting the flow value and hysteresis within zero cut-off settable range, the on-off point varies depending on the settable range.
However, please note that the setting and hysteresis of the switch output will not be changed.
To maintain the on-off point, set the value and hysteresis without the zero cut-off range.
: Zero cut-off setting does not affect the analog output.
button in function selection mode to display [F14].
Press the
button.
button to select the value of zero cut-off.
button to set.
[F14] Zero cut-off setting completed