○Error indication function
This function is to display error location and content when a problem or error has occurred.
Error
Error displayed
Over current
error
Residual
pressure
error
Pressurizing
error
System error
Description
2
The switch output load current is
80 mA or more.
2
During zero clear operation,
pressure greater than ±3.5% F.S.
is present. Note that the mode is
returned to measurement mode
automatically 1 second later. The
1
zero clear range varies by ±1%
F.S. due to variation between
individual products.
Pressure exceeding the upper
limit of the set pressure range is
1
applied.
Pressure exceeding the lower limit
of the set pressure range is
1
applied.
1
1
1
Displayed if an internal data error
has occurred.
1
1
1
1
-73-
Measures
Turn the power off and
remove the cause of the
over current.
Then supply the power
again.
Release the applied
pressure to atmospheric
pressure, and retry the
zero clear operation.
Reset applied pressure to
a level within the set
pressure range.
Turn the power off and on
again.
If the failure cannot be
solved, contact SMC.
No.PS※※-OMX1005
Error output
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable