10. Electrical connection for MULTICAL®, ULTRAFLOW®
and Pulse Transmitter
Connecting MULTICAL® and ULTRAFLOW®
ULTRAFLOW®
Blue (GND)/11A
Red (supply)/9A
Yellow (signal)/10A
ULTRAFLOW®
Blue (GND)/11A
Red (supply)/9A
Yellow (signal)/10A
Connecting via Pulse Transmitter
3,65 VDC supply
1)
Red (+)
Black (-)
1) from battery or supply module.
If long signal cables are used, please consider the installation carefully. There must
be at least 25 cm between the signal cable and all other cables due to EMC.
Pulse Transmitter
In
Out
11
11A
9
9A
10
10A
MULTICAL®
11
9
10
MULTICAL®
11
9
10
Pulse Transmitter
60
61
19