Calibration Adjustment Steps - Fluke 150 Series Calibration Manual

Insulation testers
Hide thumbs Also See for 150 Series:
Table of Contents

Advertisement

Switch Position
1000 V Insulation
Continuity
Volts
Continuity
1000 V Insulation
Continuity
Any
Must certify that this resistor is within 1.998 Ω to 2.002 Ω. This resistor should be mounted directly to the
[1]
Tester Ω and COM terminals to minimize lead resistance.
riešenia na presné meranie
Table 10. Calibration Adjustment Steps
Input Terminal
+ : COM
− : A gnd (remote test
probe pin)
+ : Volts Input
− : COM
+ : Continuity Input
− : COM
None
+: Continuity Input
−: COM
+: Battery + Terminal
−: Battery − Terminal
http://www.elso.sk
Adjustment Step
C-01
0m A, 0 Hz
C-02
15u A, 0 Hz
C-03
0.18 mA, 0 Hz
C-04
1.8 mA, 0 Hz
C-05
0.5 mA, 0 Hz
C-06
5.0 mA, 0 Hz
C-07
0.5 mA, 0 Hz
C-08
5.0 mA, 0 Hz
C-09
5.0 mA, 0 Hz
C-10
300 mA, 0 Hz
C-11
25.0 V, 0 Hz
C-12
750 V, 0 Hz
C-13
0.5 V, 0 Hz
C-14
5.0 V, 0 Hz
C-15
2.00 Ω (Use external
C-16
resistor, 0.1 %, 1 W,
50 PPM)
C-18
+ 5 V
Insulation Testers
Calibration Adjustment
Input Value
[1]