Setting The Cell Constant Manually - Xylem SI Analytics Lab 955 Operating Manual

Conductivity meter
Hide thumbs Also See for SI Analytics Lab 955:
Table of Contents

Advertisement

Lab 955
ba77144e02
12/2017
3.
Confirm the selection of (Use Cell) with <MENU/OK> or <CAL_>.
The cell constant of the last calibration is displayed.
4.
If necessary, select the suitable cell constant with <CAL>.
5.
Use <MENU/OK> to finish the calibration.
The selected calibrated cell constant is used.
5.2.5

Setting the cell constant manually

This method can be used when the cell constant of the measuring cell is known
(e.g. taken from the operating manual of the measuring cell, or from the labeling
on the measuring cell).
1.
Open the calibration menu with <CAL>.
2.
If necessary, select the calibration type (Use Cell) with <CAL>.
Use Cell is displayed.
U
S
E
C L
E
L
C
A
L
4
9
0 7
Cal
Conductivity
1/cm
21