Operation
14
4.1.2
Display
3
2
1
1 Status information
2 Measured value (with unit)
3 Measured parameter
4 Sensor symbol (calibration evaluation, cleaning interval)
5 Measured temperature (with unit)
4.1.3
Status information
Display
[AR]
Stability control (AutoRead) is active
[TP]
Temperature measurement active
[Time]
Setting of calibration interval
[ARng]
Automatic range switchover
The meter always measures with the highest possible reso-
lution.
[nLF], [Lin]
The temperature compensation is active
• [nLF]: Nonlinear temperature compensation
• [Lin]: Linear temperature compensation
[Tref20],
Reference temperature 20 °C or 25 °C
[Tref25]
[Cal]
Calibration
[LoBat]
Batteries are almost empty
TDS
Sal
8
8
8
Time
8 8
8
Tref20
Tref25
LoBat
nLF
Lin
Cal
8
S/cm
µ
*cm
MW
1/cm
8
C °
TP
°F
ARng
AR
Meaning
ba77144e02
Lab 955
4
5
12/2017