Before you install the Logic+, you need to prepare the site for installation.
Examine the location where you intend to install the cabinet. You must be
certain that the area is level and of solid construction. In addition, a dedicated
source of electrical power must be located near the installation site.
Carefully read this chapter to learn:
• Location requirements.
• Electrical power requirements.
• Exhaust requirements.
• Service utility requirements.

• Space requirements.

Refer to Appendix C: Specifications, for complete biosafety cabinet electrical
and environmental conditions, specifications and requirements.
Space Requirements
The dimensions for the Logic+ A2 biosafety cabinets are shown in Appendix B:
Dimensions.

Clearance

A minimum clearance of at least 6 inches (150 mm) is suggested on the top
and both sides of the cabinet for service.