8. Faults

8. Faults
Personnel: Skilled personnel
Protective equipment: Protective gloves
Tools: Spanner or torque wrench
EN
WARNING!
Physical injuries and damage to property and the environment
caused by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable
or toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic,
radioactive), and also with refrigeration plants and compressors, there is a
danger of physical injuries and damage to property and the environment.
Should a failure occur, aggressive media with extremely high temperature
and under high pressure or vacuum may be present at the instrument.
CAUTION!
Physical injuries and damage to property and the environment
If faults cannot be eliminated by means of the listed measures, the
instrument must be taken out of operation immediately.
For contact details, see chapter 1 "General information" or the back page of
the operating instructions.
Faults
No connection to
computer / Instrument is
not identified.
34
For these media, in addition to all standard regulations, the appropriate
existing codes or regulations must also be followed.
Wear the requisite protective equipment.
Ensure that pressure or signal is no longer present and protect against
accidental commissioning.
Contact the manufacturer.
If a return is needed, please follow the instructions given in chapter
10.2 "Return".
Causes
Cable break
No driver installed
Driver not up to date
WIKA operating instructions, models CPT2500, CPA2500 and USB-ScanSoft
Measures
Visually check the cable.
If there is any damage to the
cable, replace it with a new
one.
Install the driver
Update driver