7. Working with WIKA-Cal
7. Working with WIKA-Cal
The model CPT2500 USB pressure sensor with the model CPA2500 USB adapter can
also be operated with WIKA-Cal software.
You can find the instructions of the WIKA-Cal software on our website.
EN
7.1 WIKA-Cal configuration (also possible with the demo version)
1. In WIKA-Cal, in the Object manager, open the menu item Workstation.
⇒
The CPT2500 must first be defined as a standard and assigned to the workplace.
2. Select the communication type RS-232 via drop-down.
3. Select COM port via drop-down.
⇒
The communication is working properly if the pressure value is shown after
pressing the [Test] button.
4. Select baud rate via drop-down.
⇒
The standard baud rate is 9600 Baud.
5. Save the settings with [Save].
7.2 WIKA-Cal - Log-Template
The following options are available with the Log-Template:
New log
The [New log] function opens a new logger protocol.
Following the input of all parameters, the [Wireless address] line must be pressed in
the communication window. Select the CPT2500 used and confirm.
By pressing the [Measuring results] graphic, the logger process is started.
32
WIKA operating instructions, models CPT2500, CPA2500 and USB-ScanSoft