Performance verification
AFG3011 / 3011C
Function
Frequency
Square
2.00 MHz
2.00 MHz
Square
AFG302xB
Function
Frequency
Square
10.00 MHz
Square
10.00 MHz
AFG302xC
Function
Frequency
10.00 MHz
Square
10.00 MHz
Square
AFG305xC
Function
Frequency
Square
10.00 MHz
Square
10.00 MHz
68
Select menu
Setting
Offset
0.0 V
Channel 1 Output
On
3. Set up the Oscilloscope so the square waveform of 5 division amplitude is
displayed.
4. Verify that the rise/fall time of the square waveform on the oscilloscope at
each amplitude is equal to or less than the limit specified in the following
tables. Use 10-90% reference level for rise/fall time measurement.
5. (Two channel models only) Repeat steps 2 through 4 for the channel 2 output.
Amplitude
Offset
0.0 V
1.0 Vpp
0.0 V
10.0 Vpp
Offset
Amplitude
0.0 V
1.0 Vpp
0.0 V
10.0 Vpp
Offset
Amplitude
0.0 V
1.0 Vpp
0.0 V
10.0 Vpp
Offset
Amplitude
0.0 V
1.0 Vpp
0.0 V
10.0 Vpp
AFG3000C and AFG3000 Series Specifications and Performance Verification
Operation
TopMenu > Offset/Low (front)
On (front)
Oscilloscope
Vertical
Horizontal
200 mV/div
20 ns/div or
25 ns/div
200 mV/div
20 ns/div or
with x10
25 ns/div
attenuator
Oscilloscope
Vertical
Horizontal
200 mV/div
5 ns/div
200 mV/div
5 ns/div
with x10
attenuator
Oscilloscope
Vertical
Horizontal
200 mV/div
5 ns/div
200 mV/div
5 ns/div
with x10
attenuator
Oscilloscope
Vertical
Horizontal
200 mV/div
5 ns/div
200 mV/div
5 ns/div
with x10
attenuator
Measurement
Limit
≤ 50 ns
≤ 50 ns
Measurement
Limit
≤ 18 ns
≤ 18 ns
Measurement
Limit
≤ 9 ns
≤ 9 ns
Measurement
Limit
≤ 7 ns
≤ 7 ns