Performance verification
Harmonic Distortion Test Record (cont.)
Harmonic Distortion
(amplitude 1 V
)
p-p
reading -
reference
(dBc)
Sine 25 MHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
CH2 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
Sine 50 MHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
CH2 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
AFG310x / 310xC /
3151C / 3152C
Sine 1 MHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
CH2 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
Sine 5 MHz
CH1 Harmonic
Distortion (dBc)
40
Fundamental
= reference
2nd
0 dBc
25 MHz
50 MHz
0 dBc
0 dBc
50 MHz
100 MHz
0 dBc
0 dBc
Spectrum Analyzer reading
1 MHz
2 MHz
0 dBc
0 dBc
5 MHz
10 MHz
AFG3000C and AFG3000 Series Specifications and Performance Verification
3rd
4th
75 MHz
100 MHz
150 MHz
200 MHz
3 MHz
4 MHz
15 MHz
20 MHz
5th
Limit
Nth - reference <
-40 dBc
125 MHz
Nth - reference <
-40 dBc
Nth - reference <
-40 dBc
250 MHz
Nth - reference <
-40 dBc
Nth - reference <
-40 dBc
5 MHz
Nth - reference <
-50 dBc
Nth - reference <
-50 dBc
25 MHz