Frequency, Amplitude, DC Offset, and AC Flatness Test Record (cont.)
Frequency 50.00 MHz
Frequency 100.0 MHz
Frequency 150.00 MHz
Frequency 240.0 MHz
Harmonic Distortion Test Record
Harmonic Distortion
(amplitude 1 V
)
p-p
AFG3011 / 3011C
Sine 20 kHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
Sine 100 kHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
Sine 1 MHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
Sine 10 MHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
AFG302xB / 302xC
Sine 20 kHz
CH1 Harmonic
Distortion (dBc)
reading -
reference
(dBc)
AFG3000C and AFG3000 Series Specifications and Performance Verification
Reference - 0.50 dB
Reference - 1.00 dB
Reference - 1.00 dB
Reference - 2.00 dB
Fundamental
= reference
2nd
Spectrum Analyzer reading
20 kHz
40 kHz
0 dBc
100 kHz
200 kHz
0 dBc
1 MHz
2 MHz
0 dBc
10 MHz
20 MHz
0 dBc
Spectrum Analyzer reading
20 kHz
40 kHz
0 dBc
3rd
4th
60 kHz
80 kHz
300 kHz
400 kHz
3 MHz
4 MHz
35 MHz
40 MHz
60 kHz
80 kHz
Performance verification
dB
Reference + 0.50 dB
dB
Reference + 1.00 dB
dB
Reference + 1.00 dB
dB
Reference + 2.00 dB
5th
Limit
100 kHz
Nth - reference <
-55 dBc
500 kHz
Nth - reference <
-55 dBc
5 MHz
Nth - reference <
-45 dBc
50 MHz
Nth - reference <
-45 dBc
100 kHz
Nth - reference <
-60 dBc
37