2.3 Specifications
(26) Power supply
AC 100 to 240 V ± 10 %, 50/60 Hz, 310VA.
*1) Performance specification should be checked in the
following combination. Avoid using the non-slit septum or
the sealed plastic cap to acquire original good
reproducibility performance.
Part No.
080-3595
855-2351
080-3704
In the other cases (refer to Section 7.3), adjust the rack
parameters (refer to instruction manual for maintenance
software) according to the combination used for your
practical application, and check the reproducibility
performance before proceeding to operation.
For good reproducibility, avoid using a septum repeatedly
to prevent a cause of a hole. It causes a fluctuation of
sample concentration in sample vial.
*2) Avoid rapid changes in ambient temperature, and not
exposed to drafts such as an air conditioner. These may
influence the analysis result which reproducibility with
high precision required.
2 - 52
Part Name
1.5 mL sample vial
Septum (Si -TEF), Slit (-)