2.2 Operating Principle and Functional Description
(i)
When [wash before sample aspiration] is set.
Deliver the wash fluid1 with the wash pump, move
the needle into the rinse port, and wash the needle
outer wall (Figure 2- 26).
Wash the injection port flow path with the wash
fluid1 after the needle has moved into the injection
port (in the same procedure of Section 2.3.3 (i)).
Figure 2-26 Rinsing the Needle Outer Wall
2 - 30