Specifications; Main Unit - Hitachi ChromasterUltra Rs 6270 Instruction Manual

Table of Contents

Advertisement

2.3 Specifications

2.3.1 Main Unit

(1)
Maximum number of samples
120 (1.5 mL)
72 (4 mL) (option)
195 (1 mL) (option)
192 (96-holes microplate  2) (option)
768 (384-holes microplate  2) (option)
(2)
Sample injection method
Cut injection method
All volume injection method
Full loop injection method
(3)
Syringe capacity
70 µL (standard)
175, 700 µL (option)
(4)
Sample injection volume
0.1 to 20 µL
(5)
Loop capacity
40 µL (standard)
1, 100, 1000 µL (option)
(6)
Syringe speed
5 steps
(7)
Reproducibility of injection volume*
≤ 0.15 % RSD (in 10 L injection, Cut injection method)
≤ 0.20 % RSD (in 5 L injection, Cut injection method)
≤ 0.50 % RSD (in 1 L injection, Cut injection method)
≤ 0.20 % RSD (in 10 L injection, All volume injection method)
≤ 0.40 % RSD (in 5 L injection, All volume injection method)
≤ 0.70 % RSD (in 1 L injection, All volume injection method)