7.2 Sample Racks
b) Sample rack for thermostat (1 mL x 195)
c) Sample rack for thermostat Microplate rack (2 pcs)
v) Temperature distribution:
± 2 °C
(Target temperature 4 °C /40 °C, ambient temperature 25 °C,
relative humidity 60 %, water temperature measured in central and
4-corner sample vials (4 mL), without sample freezing)
vi) The number of the sample vial: 72 vials
vii) Applicable sample vial: 4 mL (15 mm OD x 46 mm H)
i) Applicable product: ChromasterUltra RS 6270 Autosampler
ii) Temperature setup range: 1 – 45 °C (in increments of 1 °C)
iii) Temperature control range: [Ambient temperature -21 °C] to
[Ambient temperature + 25 °C] and within the setup range (humidity
60 %, sample vial 1 mL)
iv) Temperature setup accuracy:
± 3 °C
(Target temperature 4 °C/40 °C, ambient temperature 25 °C,
relative humidity 60 %, water temperature measured in central
sample vial (1 mL), without sample freezing)
v) Temperature distribution:
±2 °C
(Target temperature 4 °C /40 °C, ambient temperature 25 °C,
relative humidity 60 %, water temperature measured in central and
4-corner sample vials (1 mL), without sample freezing)
vi) The number of the sample vial: 195 vials
vii) Applicable sample vial: 1 mL (8 mm OD x 40 mm H)
i) Applicable product: ChromasterUltra RS 6270 Autosampler
ii) Temperature setup range: 1 – 45 °C (in increments of 1 °C)
iii) Temperature control range: [Ambient temperature -21 °C] to
[Ambient temperature + 25 °C] and within the setup range (humidity
60 %, microplate)
iv) Temperature setup accuracy:
± 4 °C
(Target temperature 4 °C/40 °C, ambient temperature 25 °C, relative
humidity 60 %, water temperature measured in central and 4-corner
(each microplate), without sample freezing)
v) Temperature distribution:
± 3 °C
(Target temperature 4 °C /40 °C, ambient temperature 25 °C, relative
humidity 60 %, water temperature measured in central and 4-corner
(each microplate), without sample freezing)
vi) The number of the sample vial: 2 vials
vii) Applicable sample vial: 96, 384 holds microplate