Hitachi ChromasterUltra Rs 6270 Instruction Manual page 71

Table of Contents

Advertisement

2.2.3
(ii) When [wash before sample aspiration + fluid2
wash] is set.
Deliver the wash fluid1 with the wash pump, move
the needle into the rinse port, and wash the needle
outer wall (Figure 2- 26).
Wash the injection port flow path with the wash
fluid1 after the needle has moved into the injection
port (Figure 2- 27).
And then, wash the injection port flow paths using
the wash fluid2 (in the same procedure of Section
2.3.3 (j)) (Figure 2-28).
Figure 2-27 Washing the Injection Port Flow Path with the Wash Fluid1
2 - 31