5.3 Consumables and Spare Parts
NOTES: 1. Turn the taper side of the injection port seal to the
injection port side when attach it.
2. Looseness of the injection port may cause malfunction
of the instrument. Tighten the injection port firmly.
(8)
After replacing the injection port seal, start maintenance
software and click Maintenance - Autosampler tab - Wash for
Needle Washing to carry out injection port cleaning (wash
fluid1, 60 sec.)
NOTES: 1. After replacement of the injection port seal, make sure
liquid is not leaking through the seal.
2. If the tip of needle is deformed, the service life of port
seal may be significantly shortened. Use the needle
free from tip deformation.
5 - 24