4.6 Mounting steps version C/D

Danger:
The individual wires of the transducer cable have a voltage of
approx. 70 V during operation (see chapter "Mounting
preparations"). There is danger of electric shocks! Carry out
the mounting steps only in the complete absence of line
voltage.
Warning:
Idle condition must be provided before separating the
transducer cable. Non-observance will cause damage of the
electronics!
Proceed as follows:
à Loosen the plug connection on the socket of the previous
emitting electronics
Fig. 14: Dismounting the previous emitting electronics
1
Emitting electronics with socket
2
Plug connector
Connection piece swivelling holder or transducer tube
3
à Provide the plug connection to the socket of the new
emitting electronics
1
2
Mounting
3
15