Wi-Fi Set Up; Verifying Data Upload To Qcdm; Application Workflow; Starting The Audit - 3M Clean-Trace Luminometer LX25 User Manual

Table of Contents

Advertisement

Wi‑Fi Set up
Wi‑Fi Connection
Data is automatically uploaded to the QCDM in real-time using
the Wi-Fi function on the Clean-Trace Luminometer LX25 .
Before the Wi-Fi function can be used, the Clean-Trace
Luminometer LX25 must first be registered to the QCDM
using the USB cable connection to a workstation running the
Synchronization Manager Software .
Set up Wi‑Fi
• Tap the "hamburger menu" in the
upper right-hand corner of the
Home screen .
• Tap "Settings"
• Tap "Wi-Fi Settings"
• Turn on Wi-Fi by sliding the indicator from "OFF" to
read "ON"
• Select the appropriate Wi-Fi network and enter security
credentials if required .
Security credentials are stored in the Clean-Trace
o
Luminometer LX25 .
Authentication takes place each time the Clean-Trace
o
Luminometer LX25 is turned on and Wi-Fi is enabled .
• Tap "Done"
• Return to the "Settings" menu by tapping the Wi-Fi
network icon in the upper left hand corner of the screen .
• Exit the "Settings" screen by tapping the < button in the
upper left corner of the screen .

Verifying Data Upload to QCDM

• When connected to Wi-Fi,
test results will automatically
be uploaded to QCDM . A "Link"
icon will appear after successful
upload . An unsuccessful upload
will result in a triangle .
• If using the Synchronization Manager Software to upload
results, refer to section on "Routine Synchronization of
Results" in this document . All results will be marked with
a triangle until LX25 results are synchronized and data is
uploaded to the QCDM software .

Application Workflow

Before access to the workflow screens is available, an
organization hierarchy must be set up in QCDM, and the
Clean-Trace Luminometer LX25 must be registered to a
QCDM organization .
Workflows are application specific . There are applications
available for assessing the cleanliness of flexible endoscopes,
environmental surfaces and surgical instruments .
Starting the Audit
• Power up the Clean-Trace Luminometer LX25 .
• After power up the Clean-Trace
Luminometer LX25 will run a
device diagnostic .
• After the diagnostic is complete
a landing screen appears .
Tap "Log In"
• If the organization is part of a
multi-hospital set up, select
desired organization from the
drop-down menu then select user .
• Enter PIN
An initial user and PIN can be
o
found in the Welcome Letter
emailed to the facility when
the organization was first set
up . A User name and PIN is
established for each operator of
the Clean-Trace Luminometer
LX25 . The User can change
the assigned User and PIN
the first time logging onto the
Clean-Trace Luminometer LX25 .
9