Test 1 : Miscellaneous test
0
0
Tachometer test :
Turn the rotor by hand : the speed should appear on the 5 digit display.
Lid solenoid test :
Press on the key corresponding to program 2 to power and/or release the solenoid.
Resetting of the 5 programs :
Keep pressing on the key corresponding to program 4.
Complete program resetting on RAM :
Press simultaneously on the Δ speed programming key and the PULSE key to allow
complete resetting of the machine when it is next switched on.
- Press START to pass to the next test
- Press STOP to leave the test mode
Test 2 : Run linking
This test safety used in the factory, starts a series of preset runs after having pressed the
pulse key :
C
Y
During the run, when rotating, displays are the same as they would be outside the test
mode.
Between two runs (1 minute pause) displays are :
C
Y
- Press START key to go to the following test
- Press STOP key to exit from the test mode
Service Manual 11290099
0
0
0
T
-
1
2
0
C
L
E
-
-
-
6
C
L
E
0
0
0
2
5 digit display : speed read
3 digit display : T-1 for Test 1
2 digit display : actual temperature
Run cycles of the selected program can be run with a one
minute pause between each
Press Δ and ∇ to change the number of runs from 1 to 99
Cycles won't start if cycles number is less than 3.
Reserved for factory testing
Next run number
page 19 of 44