Siemens SITRANS FUS1010 Quick Start Manual page 39

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS FUS1010:
Table of Contents

Advertisement

5. The drop down menu lists the following sensor selections:
6. For this example, select the sensor model that appears on the sensor label then press
7. To select Sensor Size, press <Right Arrow>. Scroll to select the sensor size that matches
8. At [Sensor Mount Mode], press the <Right Arrow>. Scroll to select [Reflect] or [Direct]
9. IMPORTANT: Record Spacing Method and Number Index. This data will be used to
10. Sensors can now be mounted. Refer to mounting procedures and select the mounting
11. After sensors are mounted scroll to [Install Complete] and select [Install].
FUS1010 IP65 NEMA 4X & IP66 NEMA 7 Quick Start
Operating Instructions, 01/2013, A5E03086486-AC
– 1011 Universal
– 1011HP-T1 - Usable -40 to 120°C, recommended for Ø Temperature <40°C;
Standard.
– 1011HP-T2 - Usable -40 to 120°C, recommended for Ø Temperature >40°C - <80°C;
Named as high temperature.
– 1011HP-T3 - Usable -40 to 120°C, recommended for Ø Temperature >80°C <120°C;
special request.
– 991 Universal
Note
The meter will automatically recommend a sensor depending on the application data
that has been entered.
<ENTER>.
Select based on type.
Select based on size
After sensor is mounted select "Install.
the size indicated on the sensor label. Press <ENTER>.
mount and then press <ENTER>.
mount the sensors.
mode desired.
Connecting
3.4 Programming the Transmitter
37

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans fuh1010Sitrans fue1010

Table of Contents