Service, Maintenance And Disposal; Calibration And New Adjustment; Cleaning; Battery - Camille Bauer LINAX PQ5000MOBCL MULTI-PQ System Handbook

Table of Contents

Advertisement

9. Service, maintenance and disposal

9.1 Calibration and new adjustment

Each device is adjusted and checked before delivery. The condition as supplied to the customer is
measured and stored in electronic form.
The uncertainty of measurement devices may be altered during normal operation if, for example, the
specified ambient conditions are not met. If desired, in our factory a calibration can be performed,
including a new adjustment if necessary, to assure the accuracy of the device.

9.2 Cleaning

The operating keys should be cleaned in regular intervals. Use a dry or slightly moist cloth for this.
Damage due to detergents
Detergents may damage parts of the device. Therefore, do not use detergents.

9.3 Battery

The device contains a battery for buffering the internal clock. It cannot be changed by the user. The
replacement can be done at the factory only.
A maintenance-free capacity module is installed to bridge power failures.

9.4 Disposal

The product must be disposed in compliance with local regulations. This particularly applies to the built-in
battery and the capacity module.
Potential for Fire or Burning. Do not disassemble, crush, heat or burn the
removed batteries.
PM 1002825 000 00
System handbook PQ5000MOBCL-MultiPQ
46/60

Advertisement

Table of Contents
loading

Table of Contents