6

6 . step-By-step Instructions

A. Content
Check the contents of the equipment before use:
– one Detect
apex locator
– one measurement cable
– two connection hooks
– one user manual
B. Recharging the Battery
The Detect
is delivered with a rechargeable battery.
On the Detect
a screen indicator shows the level of the battery charge. When this is
flashing, the battery requires recharging. However, it is still functional for several treatments
before the battery shuts down.
Procedure for recharging the battery:
1. Disconnect the measurement cable
2. Connect the charger cable to the Detect
3. Connect the charger to the mains
Charging screen will appear during battery charging. When the battery image on the
screen stops blinking and turns green, the charging is complete.
ChargIng sCreen Color gUIDe
RED
LOW
ORANGE
MID
GREEN
FULL
Duration of charging: About 12 hours (24 hours after long periods without use).
Note: Detect
cannot be used while charging.
– one charger
– two lip clips
– two connection forks
C. Replacement of the rechargeable battery
The battery compartment is located at the bottom of Detect
by a screw.
1. Release the screw and remove the battery compartment cover
2. Remove the battery and disconnect the battery cable from the jack connector
3. Insert the new battery cable into the jack connector
4. Insert the battery into the battery compartment
5. Close the battery compartment and secure it with the screw
Warning: Use only the original battery pack from your supplier!
7
and the cover is secured