For models with a temperature module, plug the other end of the USB cable into the
temperature module. Insert the USB cable into the retaining clip mounted on the
temperature housing.
For models with a printer, plug the other end of the USB cable into the printer while
pulling the plastic latch away from the connector. Once the connector is seated,
push the plastic latch over the connector to secure it.
Verify that the Ethernet cable passes freely through the space between the
temperature housing and the printer board, exiting the case in the space between
the top of the temperature housing and beneath the top housing.
When installing the top housing on the chassis, line up the slots on the top housing
with the ends of the horizontal struts on the chassis to ensure proper seating.
When installing the housing top extension (75), verify that the top housing gasket is
in place. Use a 5/16-inch socket and Torque driver calibrated for 12.0 inch-pound
±1.0 inch-pound to secure the standoffs (74).
When replacing the top housing
For models without a printer, install the printer door blank as follows:
1. Remove the paper backing from the gasket on the door blank.
2. Install the door blank into the top housing, pressing to secure the door blank to the
top housing.
Remove the printer
Before you begin, remove the following as described in this manual:
Battery
Rear housing
CO2 module (extended chassis only)
SpO2 module or blanking panel
Disassembly and repair 85