OsteoSys Co., Ltd.
Model Designation : SONOST-2000
NOTE !
Try to place the system far from power generators, X-ray machines, broadcasting
stations, and transmission lines to avoid electrical noise during scanning. Otherwise,
it may lead to abnormal result. An independent circuit and a safely grounded outlets
are strongly recommended for SONOST-2000. Poor or abnormal images may occur
if the system shares a power source with other electrical or electronic equipment.
NOTE !
The classification according to standard IEC/EN 60601-I
Type of protection against electric shock : Class I
Degree of protection against electric shock : Type B
Degree of protection against harmful ingress of water : IPX 0
Methods of sterilization or disinfection : See Chapter 4.2.1
Not suitable for use in the pressure of flammable anesthetic mixtures.
4) Check followings while the device is in operation.
A) Whether the product and the patient are normal.
B) Whether the products keep distances with one another.
C) If you find any abnormality related to the product or patient, stop its
operation and check if it can cause any danger to the patient.
5) Checking points after use of product
A) Make sure to clean the product. Deactivate the product and turn the
power off in order.
B) Do not pull the power cord with an excessive force.
DOC. No. : OT01-2R7123