Model Designation : SONOST-2000
Contacting with
10
body fluids or
others
11
Measurement site
Guidance and manufacturer's declaration – electromagnetic emissions
The model SONOST-2000 is intended for use in the electromagnetic environment specified below. The customer
or the user of the model SONOST-2000 should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Need to apply ultrasound gel in between skin and probe in order to get
proper contact for path of ultrasound waves.
There are no contact between anybody fluid of human during the
N/A
examination process.
Correct position for measurement for calcaneus bone on heel as below
That point is just below area of ankle and then apply ultrasound gel with
enough amount just like below instruction. (need to apply to both side
of feet)
*According to the article Selection of the optimal skeletal site for fracture
risk prediction, the most significant relationships of BMC to fracture risk
were observed for the Oscalcis (Richard D. Wasinch, Philip D. Ross, Lance
K. Heilbrun and John M. Vogel)
*According to the article Diagnostic value of calcaneal quantitative
ultrasound in the evaluation of osteoporosis in middle-aged and elderly
A
patients, the study assessed calcaneal QUS and DXA in senile
osteoporosis diagnosis performance, as osteoporosis screening not only
has good specific degrees, and with the lumbar spine and proximal
femur DXA has good correlation, in screening for large-scale people play
the role of more convenient, economic, Therefore, calcaneal QUS can be
recommended as a pre-screening tool to determine whether DXA
screening should be performed and timely treatment should be
performed I patients with pre-existing risk of osteoporosis and patients
with mild osteoporosis to reduce the risk of fracture. (Changzhou Li,
Jifeng Sun, Li Yu)
Compliance
Group 1
Class A
Class A
Electromagnetic environment - guidance
The model SONOST-2000 uses RF energy only for its internal
functions. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
The model SONOST-2000 is suitable for use in all establishments
including domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used
for domestic purposes.