Model Designation : SONOST-2000
3.2.2.8
(1)
(2) To calibrate the system, click the "Daily Test" button from the
Daily test of SONOST-2000
NOTE! Daily Test Recommendation
The daily test should be performed within operating temperature.
(Recommended temperature: 18 ~ 27 ℃)
The phantom should be kept beside of device closely.
Daily test should be done just after turning on the device.
After turning on the device for the first time of the day, immediately
perform Daily Test immediately before warming up the device.
Turn off the device after using at closing time of the day.
Make sure to run Daily Test for the system to reduce a margin
of error and to get accurate values in times of examining a
patient's bone density. It is to raise credibility of examination
results through checking of system reliability on a daily basis.
Whenever you calibrate the system, you should use the
QC Phantom included in the accessory list.
3.1.2 Using the QC Phantom
main menu. Then the "Daily Test" screen appears as shown
below. When you click the "Daily Test Start" button in this
screen, ultrasonic waves are generated from the probe.