Two Step Blowout

During aspiration, the piston of your electronic pipette moves up (1).
1
During dispensing, the piston returns to the initial position (0). During the last
dispense of a program, it automatically moves further down (2) and blows the
0
remaining liquid from the tip (Blowout).
2
When the piston returns to the initial position (0), a small amount of air is
aspirated, provided the tip is no longer immersed in the liquid (Blowin).
Note: Manually delay the blowin by holding RUN pressed during the last dispense. Remove the
tips from the vessel and release RUN to start blowin.

Tip Spacing (VOYAGER only)

In each pipetting mode, the available tip spacing positons are displayed at the
bottom of the screen. The current position is highlighted.
At any program step press  or  to move the tip position to the next spacing.
To modify the tip spacing positions go to the "Tip Spacing" menu, accessible
from the main menu of the pipette or within every pipetting mode.
Select the positions to be changed (First, Middle, Last) and press OK. Use
 and  to move the tips to the desired position, click OK and then Save .

Equipment Disposal

Electronic pipettes must not be disposed of with unsorted municipal waste.
Dispose of electronic pipettes in accordance with the laws and regulations in your area
governing disposal of devices.
In certain regions and countries, e.g. in all EU member states, the distributor is obliged
to take back this product free of charge at the end of life. Please contact your local
distributor for more details.

Manufacturer

INTEGRA Biosciences AG
CH-7205 Zizers, Switzerland
T +41 81 286 95 30
F +41 81 286 95 33
161951_V13_QSG_Electronic_Pipettes.docx
info@integra-biosciences.com
www.integra-biosciences.com
INTEGRA Biosciences Corp.
Hudson, NH 03051, USA
T +1 603 578 5800
F +1 603 577 5529
3/7

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VIAFLO and is the answer not in the manual?

Subscribe to Our Youtube Channel

This manual is also suitable for:

Voyager

Table of Contents