Result Field; Other Messages - METREL EurotestXE MI 3102H Instruction Manual

Table of Contents

Advertisement

MI 3102H

4.1.4 Result field

4.1.5 Other messages

Hard Reset
No probe
First measurement
Last measurement
Memory full
Already saved
CHECK SUM ERROR
Fuse F1 (continuity circuit) blown or not inserted.
Noise voltage is present between H and E or S test
terminals.
Resistance of auxiliary earth electrode is higher than
100×R
. Check the auxiliary earth electrode.
E
Probe resistance is higher than 100×R
probe.
Resistances of auxiliary earth electrode and probe are
higher than 100×R
E
Measurement passed.
Measurement failed.
Measurement is aborted. Check the conditions at the
input terminal.
Instrument settings and measuremet parameters/limits
are set to initial (factory) values. For more information
refer to chapter 4.5.5. Recalling original settings.
Luxmeter sensor is turned off or disconnected from the
EurotestXE 2,5 kV instrument. Connect probe to the
instrument using RS 232 connector and turn it on.
First stored measurement results are displayed.
Last stored measurement results are displayed.
All memory locations are occupied.
Measurement results already saved.
Memory contents damaged. Contact your distributor or
manufacturer for further information.
20
Instrument operation
. Check all probes.
. Check the
E