Covers; Installing/Removing The Covers On The Support - Siemens ARCOSKOP Troubleshooting Manual

Table of Contents

Advertisement

Prerequisites
Fig. 2
Fig. 4

Covers

WARNING
NOTE

Installing/removing the covers on the support

To loosen the fastening screws for the end caps on the support arms, the rubber at the
ends must be pulled out from the slot, refer to Fig. 2 to Fig. 5. After removing the Phillips
screws, press the internally located retainer a little to the inside with a thin screwdriver
and simultaneously pull the end cap out a little. After the end cap has been loosened on
both sides it can be removed.
To remove the two-part cover of the support on the false ceiling, remove the two white
rubber plugs and unscrew the Allen screws, refer to the figure at the right above. The two
parts must be pulled apart. Take care in this case that the two internally located retaining
Siemens AG
Medical Solutions
Electrical voltage!
See Chapter 1, Safety Information.
Prior to removing the covers, switch off the line voltage supply to
the system and secure against accidental switch-on.
Switch off the compressed air brakes, turn off the compressed air
and let it escape.
SPR2-250.840.01
Rev. 01
Fig. 3
Fig. 5
Seite 7 von 16
09.05
CS PS 24
1 - 7
1
1
ARCOSKOP
System Manual

Advertisement

Table of Contents
loading

Table of Contents