5 - 38
Potentiometer
Fig. 8
Checking the programmed value
NOTE
Prerequisites
•
T
The
•
Connect the service PC to the serial interface of the ARCOSKOP.
•
Start the service program.
•
Select the "Adjustment, Calibrations..." menu.
•
Answer the query with "Yes".
•
Select "Diamentor" in the "Function Group:" combo box.
•
In the "Available Functions" window select the line "Start Diamentor Test" and click on
"Execute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No".
•
Press the >0< key (reset key for fluoroscopy time). The current value is displayed on the
monitor.
•
The tested
•
This value must match the programmed CONSTANT parameter
•
If the ->0<- key is pressed (reset key for fluoroscopic time) the test is repeated and is
displayed again on the monitor.
•
The potentiometer (Fig. 8) of the area dose product measuring device can be used to
adjust the test value
•
After the adjustment, select "Stop Diamentor Test" on the service PC and click on
"Execute".
•
Close the status window by clicking on the OK button.
•
Respond to the query "Do you want to reset the unit" by clicking "No".
ARCOSKOP
System Manual
Checking the programmed values will activate the self-test in the
electronics unit of the area dose product measuring device.
This serves to recalibrate the electronics unit.
value is already programmed (see Programming).
w
T
value is displayed on the monitor of the ARCOSKOP.
w
T
to the programmed value
w
Adjustments / Programming
T
w
SPR2-250.840.01
Rev. 01
09.05
T
. Tolerance: ± 5 %.
w
.
Seite 38 von 42
CS PS 24
5
Siemens AG
Medical Solutions