Operating software
Releasing the resetting of the load counter values
•
Select the "Adjustment / Parameters..." menu.
•
Select "SETTINGS FOR SERVICE PC" in the "Parameter Groups:" combo box.
•
Click on the "Get from Unit" button.
•
Select the parameter "ENABLE RESET TUBE VALUES".
•
In the "Value / Actual" field, select "YES".
•
Click on the "Set Value" button.
•
Click on the "Put to Unit" button.
•
Answer the query with "YES" and wait until the unit has rebooted.
•
Close the Parameters window by clicking on the "Cancel" button.
Resetting the load counters
•
Select the "Adjustment / Calibrations..." menu.
•
Answer the query with "YES".
•
Select "RESET TUBE VALUES" in the "Function Groups:" combo box.
•
Select "LOAD COUNTER" and click on "Execute".
•
The load counters are set to 0.
•
Close the Status window by clicking on the "OK" button.
•
Answer the query "Do you want to reset the Unit" with "Yes".
Returning the defective POWERPHOS incl. load counter data disk
•
Mark the load counter data disk label with:
the current date
the system part no (ARCOSKOP)
the system serial number
the POWERPHOS part number and
the POWERPHOS serial number.
•
Add the disk to the defective POWERPHOS and send it all back for repair together with
the completed tube questionnaire.