Operating software

Releasing the resetting of the load counter values

Select the "Adjustment / Parameters..." menu.
Select "SETTINGS FOR SERVICE PC" in the "Parameter Groups:" combo box.
Click on the "Get from Unit" button.
Select the parameter "ENABLE RESET TUBE VALUES".
In the "Value / Actual" field, select "YES".
Click on the "Set Value" button.
Click on the "Put to Unit" button.
Answer the query with "YES" and wait until the unit has rebooted.
Close the Parameters window by clicking on the "Cancel" button.

Resetting the load counters

Select the "Adjustment / Calibrations..." menu.
Answer the query with "YES".
Select "RESET TUBE VALUES" in the "Function Groups:" combo box.
Select "LOAD COUNTER" and click on "Execute".
The load counters are set to 0.
Close the Status window by clicking on the "OK" button.
Answer the query "Do you want to reset the Unit" with "Yes".

Returning the defective POWERPHOS incl. load counter data disk

Mark the load counter data disk label with:
the current date
the system part no (ARCOSKOP)
the system serial number
the POWERPHOS part number and
the POWERPHOS serial number.
Add the disk to the defective POWERPHOS and send it all back for repair together with
the completed tube questionnaire.
Siemens AG
Medical Solutions
SPR2-250.840.01
Rev. 01
09.05
Seite 3 von 4
CS PS 24
4 - 3
4
4
4
ARCOSKOP
System Manual