Adjustments / Programming
Calculate the maximum tube current
Tube current displayed in mA at 125 kV * 9 (R/min)
Dose rate value, measured in R/min
Multiply the tube current to be programmed by 10, since this must be entered into the
service PC in 1/10 mA steps. Round off the value to the nearest whole number.
Example:
4.4 (mA) * 9 (R/min)
11.2 (R/min)

Programming

Connect the service PC to the serial interface of the ARCOSKOP.
Select the "Adjustment, Parameters..." menu.
Select "Limitation of skin dose rate (FL)" in the "Parameter Groups" combo box.
Click on "Get from unit".
Enter the value to be programmed in "Value".
Click on "Set Value".
Select "Put to unit". Follow the menu guidance on the service PC.
Select "Save to file".
The value to be programmed has been transferred to the ARCOSKOP and temporarily
saved to disk.
NOTE

Checking

After programming, check the maximum skin dose rate again.
Siemens AG
Medical Solutions
Tube current displayed at 125 kV:
Maximum skin dose rate:
Measured dose rate value:
In 1/10 mA: 35.3, rounded off to the nearest whole
= 3.53 (mA)
number: 35
The value to be programmed is 35
You must exit the service program with "Logoff" and "Quit".
Only then will the modified parameters be stored on the disk as a
new file!
SPR2-250.840.01
Rev. 01
= tube current to be programmed
in mA
4.4 mA
9 R/min
11.2 R/min
Seite 23 von 42