Service position for the board access of the electronics box

Fig. 10
The cover latch must be closed before loosening the screws (see Fig. 10).
Remove the screws from the sides of the cover (see Fig. 10).
Be aware of the weight added by the components attached to the rear side when opening
the latch and the service cover (see 2/Fig. 11).
Fig. 11
You can now lower the generator unit. Use an extension to loosen the two screws on the
upper sides (see 3/Fig. 11).