4 - 2
Fig. 1

Shock wave head

Shock wave head covers

Switch the system off. Remove the covers on the system unit by removing the screws.
Remove the water reserve tank and drain the remaining water.
Reinstall the water reserve tank.
Switch the system on. Position the support arm in the horizontal position.
Set service switch S2 on board D3 to position 2 (service on).
Select "service" on the control unit and then "coupling circuit".
Activate the corresponding key for "empty" (S6) on the control unit until the coupling
bellow is located next to the lens.
Switch the system off.
Only the equipment with ultrasound localization:
- Unscrew the two placement parts for the ultrasound.
- Ensure that the screw lengths are correct when tightening them.
Remove the mounting screws (I/Fig. 2) for the upper cover on both sides of the shock
head.
Push the buttons in the direction of the arrow on both sides of the iso-center phantom
(Fig. 2) and at the same time flip the iso-center phantom back (Fig. 1).
Detach the upper cover, removing it toward the front.
Push the buttons on both sides of the iso-center phantom in the direction of the arrow
(Fig. 1) and at the same time flip the iso-center phantom up (Fig. 2).
If required, remove the 3 cover screws from both lower covers and remove the covers.
Reattach the covers in the reverse order after completion of all work.
MODULARIS Uro Plus
LITHOSTAR MODULARIS
Fig. 2
SPL1-130.840.02
Page 2 of 34
Rev. 03
04.05
CS PS 24
I
4
4
Siemens AG
Medical Solutions