Fault
7 blink impulses
8 blink impulses
9 blink impulses
10 blink impulses
Cause
A system or joystick error.
The operating module or the elec-
tronic is defective or a system error
has occurred.
Fault on the magnetic brakes of
the motor.
The battery voltage is too high
(downhill driving).
Remedy
Do not touch the joystick during
the initiation phase. – Switch the
electronic travelling wheelchair
off and on again.
Check cables and connecting
plugs. – Switch the electronic
travelling wheelchair off and on
again.
Put the drive/push mode selec-
tion lever into the drive mode
position.
Only drive downhill very slowly
and if necessary check the bat-
tery contacts.