Error Messages 

14. ERROR MESSAGES

The instrument shows clear warning messages (bottom of the screen) when erroneous conditions appear
or when measured values are outside the expected range.
The information below provides an explanation of the errors and warnings, and recommended action to be taken.
Displayed Message
"Temperature under/over range"
"Under/over compensation range"
"EC under/over range"
"Broken electrode"
"EC out of calibration range"
"Factory calibration expired"

15. ACCESSORIES

CONDUCTIVITY SOLUTIONS
HI7030M or HI7030L
HI7031M or HI7031L
HI7033M or HI7033L
HI7034M or HI7034L
HI7035M or HI7035L
HI7037M or HI7037L
HI7039M or HI7039L
OTHER ACCESSORIES
HI740036P
HI740037P
HI764060
HI900946
HI900947
HI920016
Explanation & Recommended Action
Temperature outside specified range.
Replace probe if necessary.
During conductivity calibration, the temperature is under/over the
conductivity calibration solution compensation limit.
Outside range in the corresponding scale.
Ensure the vent holes are completely submerged and the sample is within
specified range.
The meter fails to calibrate or gives faulty readings.
Replace the probe.
Displayed when the measured value is outside calibration range.
Contact the Hanna technical support for the periodic factory calibration.
12880 μS/cm standard solution, 250 or 500 mL
1413 μS/cm standard solution, 230 or 500 mL
84 μS/cm standard solution, 230 or 500 mL
80000 μS/cm standard solution, 250 or 500 mL
111800 μS/cm standard solution, 230 or 500 mL
100 % NaCl seawater salinity standard solution, 250 or 500 mL
5000 μS/cm standard solution, 250 or 500 mL
100 mL beaker (10 pcs.)
20 mL beaker (10 pcs.)
Electrode holder
115 Vac to24 Vdc power adapter, US plug
230 Vac to 24 Vdc power adapter, European plug
USB type A to C cable