Refilling The Electrode; Storage; Warning & Error Messages; Abbreviations - Hanna Instruments HALO 2 HI9810392 Instruction Manual

Wireless ph tester for chocolate
Table of Contents

Advertisement

Refilling the Electrode

• To remove the electrode sleeve, carefully rotate and slide it off (A). Set it to one side.
Note: Handle the probe with care, the pH electrode stem is made of glass.
• Rinse off any traces of electrolyte gel from sleeve. Soak the electrode tip in
General purpose cleaning solution (or HI700661, HI700663, HI700664) for
20 minutes. Rinse with distilled or deionized water.
• Refill the reference well with
• Replace the sleeve. Ensure the O-ring is fixed inside the electrode. Rinse off excess gel
with distilled or deionized water.
• Shake the probe down gently to eliminate any trapped air bubbles.
• Soak the probe in
HI70300

Storage

To ensure a quick response, the glass tip and junction should be kept hydrated.
When not in use, add a few drops of storage solution to the storage cap. If storage solution is not available, pH 4.01 or
pH 7.01 buffer can be used.
Note: Do not store the electrode in distilled or deionized water.
11. WARNING & ERROR MESSAGES
----
WRNG
12. 0 2
23. 7
8. 0 2
!
23. 7
BAtt
DEAD

12. ABBREVIATIONS

ATC
Automatic Temperature Compensation
GLP
Good Laboratory Practice
MTC
Manual Temperature Compensation
NIST
National Institute of Standards and Technology
HI9071
Gelled bridge electrolyte (B).
Electrode storage solution for a minimum of 30 minutes before calibrating.
"---- WRNG" displayed during calibration.
Invalid buffer.
Check the buffer value and use fresh buffer.
Measured value displayed blinking.
Measured value is out of electrode range.
Clean the electrode to improve condition.
Battery icon (
) displayed blinking.
!
Battery is below 10 %, replace the battery.
"BAtt DEAD" and tester turns off.
Depleted battery, replace the battery.
A
HI700601
B
9