Service Requirements - Ametek LAND MILLWATCH User Manual

Table of Contents

Advertisement

Single Stream
3

Service Requirements

Service
1
Calibration gas*
2
Instrument air*
3
Condensate removal
4
Electrical power supply
5
Cooling air
6
Sample line connection
7
Exhaust
Must be kept at atmospheric
pressure.
Do not allow back pressure
8
Cable glands
9
Calibration gas to probe
10 Blowback air
11 Blowback signal
* these MUST be set to the same pressure
+ electrical power supply connected to terminal rail, where fitted
User Guide
Requirement
1 to 2 bar (15 to 30 psi)
CO in Nitrogen
20 litres (0.7 ft
)
3
1 to 2 bar (15 to 30 psi)
dry, regulated, clean, oil
free compressed or bottled
air
Optional
83 to 132 V or
+
165 to 264 V a.c.
50 to 60 Hz 300 W
4 bar (60 psi) minimum
8 bar (120 psi maximum)
Clean. Oil-free
15 m (50 ft) maximum
nylon/antifreeze
Optional
Customer's connections
15 m (50 ft) maximum
nylon/antifreeze
4 to 8 bar (60 to 120 psi)
24 V dc, powered by
analyzer
Millwatch / Silowatch
Size/Specification
4 l/min (0.15 ft
/min) during
3
calibration
1/4 in outside diameter PTFE,
nylon or stainless steel tube
1/4 in outside diameter PTFE,
nylon or stainless steel tube
1/2 in NPT - 14 female route
to collection device
Cable 3 way
Core size 24/0.2 minimum
300 l/min (10 ft
/min)
3
1/4 in outside diameter PTFE,
nylon or stainless steel tube
Outside diameter 1/4 in
Sample flow rate is
1.5 l/min (0.05 ft/min)
3/16 in barbed fitting for
4.7 mm inside diameter
tubing. Vent to safe area.
Cable diameter 7 to 10.5 mm
(0.58 to 0.41 in)
1/4 in outside diameter PTFE,
nylon or stainless steel tube
1/4 in outside diameter PTFE,
nylon or stainless steel tube
Cable 2 way
Core size 24/0.2 minimum
3