17.6 Set the Automatic Calibration Times and
Calibration Gas Concentrations
Checking and entering the calibration gas concentration is essential. The calibration
gas concentration is shown on the certificate, which accompanies the calibration
gas cylinder.
1)
Press the CAL button.
2)
Use the arrows to change the displayed value to the entry code.
3)
Press CAL again.
4)
Select AUTOCAL and press CAL.
The calibration menu is now displayed.
5)
To change the entries, press SETUP, change the value with the arrow keys and
press SETUP again.
TYPE
CAL INTERVAL
SETTLE TIME
SPAN GAS
REPORT DELAY
CLOP
CAL METHOD
MAINT RELAY
TIME OF NEXT CAL?
24
MANUAL
See below
TIMED TRIGGER Calibration is set by TIME TO NEXT CAL and CAL
INTERVAL values
EXT TRIGGER
Use an external calibration switch
OFF
Calibration is disabled
AIR SOLENOID AND GAS1 SPAN for LAND use only
The time between calibrations in hours (0 to 255 hours)
Time allowed for calibration gas to reach a steady state
0 to 4000 ppm. Given on the calibration gas certificate
The time required for the report relay to remain closed in order for the
recording or other device to acquire information. Default 20 seconds.
TRACK or HOLD (outputs during a calibration)
2 POINT
CAL IN
PROGRESS
MAINTENANCE
DAY hour:minute
Use the arrow keys to change the day and then press SETUP, repeat
for hour and minute
ppm or mg/Nm
3
mg/Nm
= ppm 1.25
3
(STD = 0° C and 101.3 kPa)