Maximum Deviations; Tube Housing Assemblies; Siemens Ag; Spr2-330.662.01 - Siemens ARCADIS Avantic Maintenance Instruction

Hide thumbs Also See for ARCADIS Avantic:
Table of Contents

Advertisement

6 - 2

Maximum deviations

Fluoroscopy
Tube voltage:
Tube current:
Scene timer:
Basic measurement used
Peak tube potential:
Tube current:

Tube Housing Assemblies

Maximum rated peak tube potential:
Leakage technique factors:
Minimum filtration permanently in the useful beam:
obtained at:
Minimum filtration permanently in the useful beam:
obtained at:
Cooling curves
Tube housing cooling curve:
Tube rating charts:
Definitions of the measurement bases of technique factors
Peak tube potential:
Measurement of kVp at radiography and at fluoroscopy, using PTW-Nomex (2 filter
method).
Tube current:
As measured in the mA-measuring pins D1.X39-D1.X40 (x97 opened) in the tube current
circuit near ground with an mA-meter (FLUKE 8060 A or equivalent), minus the current
flowing through the built-in high-voltage divider of 0.0025 mA/kV
ARCADIS Avantic
± 10 %
± 10 % ± 0.1 mA
0 to 5 min. tolerance 1 digit (6 sec)
(Tested IEC601/ 2 / 7)
DIADOS (2 filter method)
by using RMS-multimeter
See Operator's Manual
See Operator's Manual

SPR2-330.662.01

Rev. 02
05.06
Technical Data
125 kV
p
125 kV
/4.3 mA/550 W
p
125 kV
/7.8 mA/1000 W
p
> 3,1 mm Al equivalent
80 kV
p
> 4,5 mm AL equivalent
125 kV
p
.
p
Page 2 of 2
CS PS SP
6
6
6
6
6

Siemens AG

Medical Solutions