Notes on trouble-shooting
Checking the maximum main inverter frequency
•
POLYMOBIL OFF
•
Open bridge "S1" on D920 (solder).
•
Connect oscilloscope on D910 to MP "REG" and "GND".
•
POLYMOBIL ON
•
Trigger exposure.
NOTICE
•
Adjust the maximum main inverter frequency with potentiometer P2 to D910
(X028I-91E).
•
POLYMOBIL OFF
•
Connect jumper "S1" on D920 (solder).
Siemens AG
Medical Solutions
- Err 15 is displayed.
- No radiation.
100
F
= 10 kHz
max
µ
T
= 100
s Tolerance: + 0
min
µ
µ
s / - 5
s
RXB8-115.061.02
Rev. 01
10.95
Page 9 of 24
CS PS 24
4 - 9
4
POLYMOBIL III