Checking The Maximum Main Inverter Frequency - Siemens POLYMOBIL III Service Instructions Manual

Hide thumbs Also See for POLYMOBIL III:
Table of Contents

Advertisement

Notes on trouble-shooting

Checking the maximum main inverter frequency

POLYMOBIL OFF
Open bridge "S1" on D920 (solder).
Connect oscilloscope on D910 to MP "REG" and "GND".
POLYMOBIL ON
Trigger exposure.
NOTICE
Adjust the maximum main inverter frequency with potentiometer P2 to D910
(X028I-91E).
POLYMOBIL OFF
Connect jumper "S1" on D920 (solder).
Siemens AG
Medical Solutions
- Err 15 is displayed.
- No radiation.
100
F
= 10 kHz
max
µ
T
= 100
s Tolerance: + 0
min
µ
µ
s / - 5
s
RXB8-115.061.02
Rev. 01
10.95
Page 9 of 24
CS PS 24
4 - 9
4
POLYMOBIL III